ULVAC oil rotary vane vacuum pump schematic diagram

ULVAC Oil Rotary Vacuum Pumps

ULVAC oil rotary vacuum pumps back turbomolecular pumps, serve as standalone rough and medium vacuum sources, and are the first-stage pump in oil diffusion pump systems. The range covers single-stage and two-stage rotary vane pumps (PVD series), belt-drive oil-sealed pumps (PKS series), direct-drive two-stage pumps (VD and VS series), hermetic-drive pumps (GHD), and high-displacement pumps (GLD). All models use oil to seal rotor-stator clearances, lubricate vane tips and bearings, and remove heat. A gas ballast valve is standard on most models to handle condensable vapours.

Part Numbers

Product Families

SeriesTypeSpeed RangeNotes
PVD-180, PVD-360Single-stage rotary vane, direct drive180 to 360 L/minCompact backing pump for TMP, MBP, and diffusion pump systems
PKS-016, PKS-030, PKS-070BBelt-drive rotary vane, two-stage16 to 70 m3/hBelt drive allows motor replacement without internal oil contact
VD-30C, VD-60C, VD-90CDirect-drive rotary vane, two-stage30 to 90 m3/hMedium duty; suitable for process gas loads
VS-1501, VS-2401Direct-drive rotary vane, single-stage1500 to 2400 L/minHigh throughput for large chamber roughing and process pumping
GHD-031Hermetic drive, oil-sealed31 m3/hHermetically sealed motor for flammable or aggressive process gases
GLD-137COil-sealed, large displacement137 m3/hFor high gas load process applications

Oil and Lubrication

Oil is essential to pump operation: it seals the rotor-stator clearance, lubricates the vane tips and bearings, and removes heat. Use ULVAC Ultra Oil or an equivalent fully synthetic vacuum pump oil of the correct viscosity. Do not use general-purpose mineral oil — insufficient vapour pressure properties will degrade ultimate pressure performance.

  • Check oil level via sight glass before each shift — oil level must be between MIN and MAX marks with pump at rest
  • Oil colour: pale yellow or clear is normal; dark brown or black indicates thermal stress; grey or white milky appearance indicates water contamination — change oil immediately
  • Oil change interval: typically 2000 operating hours or annually — shorten significantly for reactive, corrosive, or condensable process gases
  • Gas ballast: open when pumping condensable vapours (solvents, water vapour) to prevent condensation in the oil; close gas ballast for dry clean gases to improve ultimate pressure
  • Always fit an oil mist separator on the exhaust — prevents environmental contamination and recovers oil carried out as mist

Installation Requirements

  • Mount on a level, rigid surface — inclined mounting or vibrating structures are not acceptable
  • Fit a solenoid inlet valve that closes automatically on pump stop — prevents oil suck-back into the process chamber on sudden shutdown
  • Exhaust: route to scrubber or ventilated area; fit oil mist separator at exhaust port; never discharge to a closed room
  • PKS belt-drive models: check belt tension at installation; belt guard must be fitted before operation
  • Verify motor rotation direction before first run — wrong rotation causes oil to spray from exhaust port
  • Solenoid valve options for PVD-B series available in voltages from AC200V to AC480V — confirm supply voltage when ordering

Pre-Commissioning Checks

  • Verify oil level is between MIN and MAX marks
  • Confirm motor rotation direction matches rotation arrow on pump body
  • Verify exhaust is connected and routed safely
  • Blank off inlet with gauge to check no-load performance before connecting to system
  • Confirm electrical supply voltage and phase match motor nameplate
  • PKS models: check belt condition and tension

Commissioning Sequence

  1. Open gas ballast valve
  2. Start pump with blank-off plate on inlet; run 5 minutes to warm up
  3. Connect calibrated vacuum gauge to inlet; record ultimate pressure with blank-off; compare to specification
  4. Close gas ballast valve; wait 5 minutes; record improved ultimate pressure (two-stage models should reach specification)
  5. Connect pump to vacuum system and open inlet valve
  6. Monitor pump body temperature — steady state is typically 60 to 80 degrees C
  7. After 30 minutes of operation: re-check oil level with pump warm; top up if necessary
  8. Record commissioning data: date, pump model, serial number, ultimate pressure blank-off, system base pressure, oil type used

Preventive Maintenance

TaskIntervalNotes
Oil level checkBefore each shiftCheck sight glass; top up with same oil type only
Oil condition visual checkWeeklyColour, clarity, any milkiness indicating water contamination
Oil change2000 hours or annuallyShorten for corrosive or wet process gases; drain warm oil for complete removal
Oil mist trap serviceAt each oil changeReplace or clean coalescing element; drain collected oil
Inlet filter inspectionAt each oil changeReplace element if contaminated; blocked filter reduces pumping speed
Belt inspection (PKS series)Every 6 monthsCheck tension and condition; replace if cracked or glazed
Exhaust valve inspectionAnnuallyCheck for cracks, debris, or carbon buildup; replace if necessary
Shaft seal inspectionAnnuallyCheck for oil leak at drive end; replace shaft seal if dripping or wet
Ultimate pressure testAnnually or after oil changeRecord and compare to baseline commissioning value

Condition-Based Maintenance Criteria

  • Change oil immediately if oil is grey/white (water contamination), black (thermal breakdown), or contains visible particulate
  • Investigate shaft seal immediately if oil drips are found under the pump at the drive end
  • Investigate exhaust valve if: pulsing noise on exhaust, oil spraying from mist trap excessively, or ultimate pressure significantly worse than specification
  • Schedule overhaul if ultimate pressure is more than 10 times worse than specification after confirmed fresh oil change and confirmed inlet valve seating

Fault Diagnosis

SymptomProbable CauseDiagnostic StepAction
Poor ultimate pressureOil contaminated; exhaust valve damaged; shaft seal worn; inlet valve leakingChange oil first; test with blank-off plate on inlet; leak test inlet valveChange oil; replace exhaust valve; replace shaft seal as indicated
Excessive oil consumptionOil mist trap saturated; oil overfilled; high process gas loadCheck mist trap condition and oil levelService mist trap; verify correct fill level; check gas ballast operation
Oil in process chamberInlet solenoid valve missing or not closing on stop; reverse rotationVerify inlet solenoid closes on pump stop; check rotation arrow on pump bodyFit or service inlet solenoid valve; correct motor wiring for rotation direction
Abnormal knocking or rattling noiseDebris in pump; vane broken; bearing wornStop pump; strain oil through fine filter; check for metal particlesInternal inspection and workshop service required
Pump overheating above 100 degrees CHigh ambient temperature; cooling fins blocked; wrong oil gradeCheck ambient temperature and cooling fin conditionImprove ventilation; clean fins; verify correct oil grade
Motor trips on overloadPump seized or very stiff; wrong supply voltageCheck pump rotates freely by hand with power off before re-startingInvestigate seize cause; verify electrical supply
Shaft seal leak at drive endShaft seal worn or hardened; excessive exhaust back-pressureConfirm shaft seal is source of leak not bearing coverReplace shaft seal; investigate back-pressure cause

Service Scope

Remote support: oil condition assessment by photo, performance data review (ultimate pressure trend), gas ballast setting advice, oil type verification, rotation direction check by video call.

On-site service: shaft seal replacement, exhaust valve replacement, internal inspection, rotor-stator clearance check, oil flush and change, full performance test post-service.

Workshop overhaul: complete disassembly, clearance measurement to specification, vane replacement, rotor and stator reconditioning or replacement, full performance test before return.

Operator and Maintenance Training

Operator training: daily oil level and condition checks, gas ballast usage, correct startup and shutdown sequences, recognising abnormal noise and oil condition, log-keeping. Maintenance training: oil change procedure, exhaust valve inspection and replacement, shaft seal replacement, oil mist trap service, belt inspection and replacement (PKS series), and performance testing. Contact mail@ulvacheliot.com — provide: pump models, number of attendees, site location, preferred dates.

Service Enquiry

Email mail@ulvacheliot.com — provide: pump model, part number, serial number, oil type in use, operating hours, process gas, symptom description, and photos of oil condition and nameplate.

Independent HELIOT Equipment Service

Independent third-party maintenance, commissioning and training provider for ULVAC HELIOT helium mass spectrometer leak detectors. We are not the manufacturer and do not sell new ULVAC equipment. Worldwide on-site support.

mail@ulvacheliot.com


Independent service provider notice: This website is operated by an independent, third-party technical service organisation. We provide on-site and remote maintenance, commissioning, repair and training services for existing ULVAC HELIOT equipment already owned or procured by our customers. We are not ULVAC Corporation, not a subsidiary, dealer, distributor, or authorised representative of ULVAC Corporation, and we do not sell new ULVAC equipment. ULVAC® and HELIOT® are registered trademarks of ULVAC Corporation (Japan), used here solely in a descriptive, nominative sense to identify the equipment we service. | Full Trademark Disclaimer | Contact